Required Update — Version 3 Program Fix
This update is required for all FirstCall™ Medical Gas Alarm Panels and Manifolds shipped between 11/22/2025 and 4/14/2026. Follow all steps below in order. Products outside this date range only need the firmware update.
1
Format USB Drive to FAT32
Prepare the USB flash drive before downloading files
  • 1
    Insert the USB thumb drive into your Windows PC.
  • 2
    Open File Explorer, select the USB drive, then open the Drive Tools tab in the ribbon.
  • 3
    Click Format.
  • 4
    Confirm the File System is set to FAT32 (Default). Do not change any other settings.
  • 5
    Click Start and confirm when prompted. The drive will be ready once formatting completes.
2
Download Program File for Your Product
Select the correct file below and save it to the formatted USB drive

Manifolds

Gas Cylinder Medical Manifold
Gas Cylinder Medical Manifolds Program ⬇ Download
Gas Liquid Medical Manifolds
Gas Liquid Medical Manifold Program ⬇ Download
Gas Liquid Lab Manifold Program ⬇ Download
Gas Liquid Medical Manifold — Original Sensors Program ⬇ Download

Area Alarm Panels

Area 7″ Alarm Programs
Area 7″ Alarm Program — 1 to 4 Gas ⬇ Download
Area 7″ Alarm Program — 5 to 8 Gas ⬇ Download
Area 10″ Alarm Programs
Area 10″ Alarm Program — 1 to 4 Gas ⬇ Download
Area 10″ Alarm Program — 5 to 8 Gas ⬇ Download

Combo Alarm Panels

Combo 7″ Alarm — 1 to 4 Gas
Combo 7″ — 1 to 4 Gas, 23 Inputs ⬇ Download
Combo 7″ — 1 to 4 Gas, 39 Inputs ⬇ Download
Combo 7″ — 1 to 4 Gas, 55 Inputs ⬇ Download
Combo 7″ Alarm — 5 to 6 Gas
Combo 7″ — 5 to 6 Gas, 23 Inputs ⬇ Download
Combo 7″ — 5 to 6 Gas, 39 Inputs ⬇ Download
Combo 7″ — 5 to 6 Gas, 55 Inputs ⬇ Download
Combo 10″ Alarm — 1 to 4 Gas
Combo 10″ — 1 to 4 Gas, 23 Inputs ⬇ Download
Combo 10″ — 1 to 4 Gas, 39 Inputs ⬇ Download
Combo 10″ — 1 to 4 Gas, 55 Inputs ⬇ Download
Combo 10″ Alarm — 5 to 6 Gas
Combo 10″ — 5 to 6 Gas, 23 Inputs ⬇ Download
Combo 10″ — 5 to 6 Gas, 39 Inputs ⬇ Download
Combo 10″ — 5 to 6 Gas, 55 Inputs ⬇ Download

Master Alarm Panels (MAP)

MAP 7″ Alarm Programs
MAP 7″ — 23 Inputs ⬇ Download
MAP 7″ — 39 Inputs ⬇ Download
MAP 7″ — 55 Inputs ⬇ Download
MAP 10″ Alarm Programs
MAP 10″ — 23 Inputs ⬇ Download
MAP 10″ — 39 Inputs ⬇ Download
MAP 10″ — 55 Inputs ⬇ Download

Firmware (Non-Affected Devices)

All Product Firmware — Unistream v12.25
Required for all products shipped before 11/22/2025. Install this instead of the program files above.
⬇ Download
3
Install Program from USB
Load the downloaded file onto the panel via USB
  • 1
    Go to the panel and open the enclosure door.
  • 2
    Insert the USB stick into the USB slot on the side of the PLC/HMI unit (the rectangular port inside the panel).
  • 3
    Power on the panel.
  • 4
    Press and hold the upper-right corner of the touchscreen until the system menu appears, then tap Login.
  • 5
    Enter the login credentials below. The login is case and spacing sensitive — use the eye icon in the password field to verify your entry before submitting.
    Username User
    Password Pass123
  • 6
    After logging in, press and hold the upper-right corner again and select UniApps.
  • 7
    Navigate to:
  • 8
    On the Upgrades screen, press Import Application, then press Import to confirm.
  • 9
    The system will reboot automatically. Once it restarts, the program update is complete.
4
Safely Remove USB Flash Drive
Always use the Unitronics safe-remove procedure — do not pull the drive directly
  • 1
    Log into UniApps using the Level 4 username and password. This will bring you to the System screen.
  • 2
    Select Control from the System screen icons.
  • 3
    Tap the Safe Remove tab at the top of the screen.
  • 4
    Select Safely Remove DOK. Once confirmed, you may physically remove the USB drive from the panel.
    ⚠ Always use this procedure before removing the USB drive to prevent file system corruption on the PLC.